Hearing test

Accept PCH and ensure initials were made Enter COSI from PCH Sign PCH Perform test

Pure tones [^1]

Pure tones all frequencies Start at 30, no response jump to 50, no response jump up 20db

UCLs

500, 1k, 2k, 3k, 4k Start at 70db

Bone Conduction

500, 1k, 2k, 4k

Masking Rules for Bone Conduction

Deliver narrowband masking noise to the NTE when the difference between the measured AC and BC thresholds of the TE is 15 dB or greater

Initial Amount of Masking in NTE = AC_N$$_T$$_E + 10 dB + OE

Initial Amount of Masking in NTE for 500Hz = AC_N$$_T$$_E + 10 dB + 15 dB = AC_N$$_T$$_E +25 dB Initial Amount of Masking in NTE for 1000Hz = AC_N$$_T$$_E + 10 dB + 10 dB = AC_N$$_T$$_E +20 dB Initial Amount of Masking in NTE for 2000Hz = AC_N$$_T$$_E + 10 dB + 0 dB = AC_N$$_T$$_E +15 dB Initial Amount of Masking in NTE for 4000Hz = AC_N$$_T$$_E + 10 dB + 0 dB = AC_N$$_T$$_E +15 dB

Occlusion Effect

Table 8-3. Occlusion Effect

textbook pg 153

Frequency (Hz)25050075010001500200030004000
Occlusion Effect (dB)151515*100000
    • = estimated value

Speech Testing

MCL

MCL script “We are going to measure your most comfortable listening level. As I speak point your finger up if you want the volume louder and down when you want it softer. Please tell me when we reach the loudness level you prefer. Do you have any questions?”

SRT (Speech Recognition Threshold)

textbook pg 167 Start spondee presentation at 30 dB HL, no response jump to 50, no response jump up 20dB

SRT is defined as the lowest level at which at least 50% of the responses are correct, with a minimum of three correct responses at that intensity

SRT should be within 10 dB of the pure tone average (PTA) unless the audiogram exhibits an unusual configuration, such as precipitous or reverse slope

Masking Rules for SRT

textbook pg 170 When masking is needed in the NTE, the initial level of masking is SRT_N$$_T$$_E + 10 dB

Word Recognition (WRS)

textbook pg 172 Never less than 70 dB? Starting volume depends on patient’s threshold at 2000 Hz:

Presentation Levels Based on 2000 Hz Threshold

textbook pg 173

Patient/Client Threshold at 2000 HzPresent Words at (dB SL)
Below 50 dB HL25 dB SL
50–55 dB HL20 dB SL
60–65 dB HL15 dB SL
70–75 dB HL10 dB SL

If a patient has a 2000 Hz threshold of >75 dB HL, there are two options for determining the presentation intensity for word recognition testing:

  1. follow the recommendations for patients with a 2000 Hz hearing level of 70-75 dB HL
  2. present the word recognition list at the 2000 Hz loudness discomfort level (LDL) minus 5 dB HL

Sign printed audiogram Create COSI using PCH Sell hearing aids Make fitting appointment Order hearing aids/place in stock clipboard Print order form and place in new orders folder

Scan PCH Scan audiogram

[^1] pg 127